Physical specifications
Screen: 5.4'' Display Screen
Suitable Adult & Child
Mode: pneumatically powered and electrically controlled system
Working Mode: Closed; Semi-Closed; Semi-Open
Circuit Breathing circuit integrated standards
Tube: 4Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min
Trolley: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Drawer unit ONE drawer totally extractable
BR-AM04 Anesthesia Machine
Technical Specifications:
Gas pressure: Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ; NO2: 0.32 MPa to 0.6 MPa.
safety valve ≦12.5 kPa
Respiratory Rate 1~99bpm
Flow meter O2:0.1~10L/Min, N2O:0.1~10L/Min
oxygen concentration in the mixed gas N2O/O2 > 21%
Rapid Oxygen Flush: 25~75 L/min
modes of ventilation IPPV, SIPPV, IMV, SIMV, MANUAL
I/E Ratio: 4:1 ~ 1:10
Tidal Volume 0~1500ML
Inspiratory trigger pressure -0.1kPa~1.0kPa
Inspiratory Plateau: 0~1s
Maximum safety pressure: ≤ 12.5 kPa
Pressure limit range: 0 ~ 6.0 kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicating: Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa
±0.2 kPa
Power Supply Alarm "Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s"
The airway pressure continues to be higher than 15 hPa?±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.