Features: 
1. 4.3” color TFT LCD screen (LED backlight), with resolution 272×480
2. Three working modes: Rate/Time/Volume mode
3. Drug library with 210 kinds of medicine 
4. Storage 1500 history records 
5. Heating function, suitable for infusing in winter or there is requirement for the fluid medication temperature
6. Support multi language display
7. A variety of visible and audible alarm, make infusion more safe

Standard:Drug library, History record, Heating function, Drip detector ,Remote control

Optional:SpO2,NIBP

Why we add NIBP &SpO2 to an infusion pump?
NIBP & SpO2 are the most frequently used vital sign parameters, we  make them work as one, to help doctors reduce cost and save space  which is already very limited in the hospital room.

R-IF03 Single-channel Infusion Pump, remote control

£0.00Price
  • Technical Specification:

    Infusion rate range
    Adjustable drop rate range: 1 /min~400 drops/min 
    (step: 1 drop /min)
    Adjustable flow rate range: 1ml/h~1200ml/h
    (step 0.1ml/h)
    Purge rate: 800ml/h
    KVO rate: 1ml/h (less than 100ml/h)
    2ml/h (100~300ml/h)
    3ml/h (greater than 300ml/h)
    Infusion volume range: 1ml~9999.9ml (step: 0.1 ml)
    Maximum accumulated volume: 9999.9ml (step: 0.1 ml)
    Infusion Time range: 1min~9999min (step: 1 min)

    Infusion accuracy: Within ±3% (using calibrated high quality IV set)
    Mechanical precision: within ±2%

    Occlusion alarm threshold:
    High: 800 mmHg ± 200 mmHg (106.7kPa±26.7kPa)
    Medium: 500 mmHg ± 100 mmHg (66.7kPa±13.3kPa)
    Low: 300 mmHg ± 100 mmHg (40.7kPa±13.3kPa)
    Alarm: Infusion Complete, Empty, Faulty Signal, Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, Idle.

    NIBP
    Range
    Average deviation: ≤ ±5mmHg
    Standard deviation:  ≤ 8mmHg
    Adult
    SYS: 40mmHg~280mmHg
    MEAN: 20mmHg~240mmHg
    DIA: 10mmHg~220mmHg
    Pediatric
    SYS: 40mmHg~220mmHg
    MEAN: 20mmHg~170mmHg
    DIA: 10mmHg~160mmHg
    Neonate
    SYS: 40mmHg~135mmHg
    MEAN: 20mmHg~110mmHg
    DIA: 10mmHg~100mmHg
    Static pressure
    Range: 0mmHg~300mmHg
    Deviation: 3mmHg

    Resolution: 1mmHg

    NIBP measurement mode
    Manual
    Cycle (automatic): 
    Optional interval: 1min, 2min, 3min,4min, 5min, 10min, 
    15min, 30min, 60min, 90min
    Fast: continuous NIBP measurement, interval is 5min

    NIBP alarm
    High pressure alarm: >90mmHg-160mmHg
    Low pressure alarm: <50mmHg-90mmHg
    Accuracy: ±5%

    Overpressure protection
    Adult: 280 mmHg ±10 mmHg
    Pediatric: 237 mmHg ±3 mmHg
    Neonate: 147 mmHg ±3 mmHg

    SpO2
    Range: 0%~100%
    Resolution: 1%
    Accuracy: 70% to 100%, precision is 2%
    0% to 69%, undefined

    SpO2 alarm
    Upper limit: 100%
    Lower limit: 90%
    Alarm value accuracy: 3%
    Alarm time accuracy: 2s

    Pulse rate
    Range: 20bpm~300bpm
    Resolution: 1bpm
    Accuracy: 3bpm

    Pulse rate alarm
    Upper limit: 120bpm
    Lower limit: 50bpm (default)
    Alarm value accuracy: 5%
    Alarm time accuracy: 2s

    NIBP Cuff requirement
    Human body contact materials: Waterproof cloth
    Size: 515mm(L)×145mm(W), allowed deviation: ±5%

    SpO2 sensor
    Type: Finger clip type
    Wire length: 3000mm, allowed deviation: ±5%
    Human body contact materials: Silicone

    Size: 188 mm(L)×198 mm(W)×228 mm(H)
    Net weight: 2.5kg