Introduction:

The BR-MV14 ventilator is gas driven by medical compressed oxygen and compressed air. In the inspiratory phase, two streams of compressed gas (compressed oxygen and compressed air) flow into a high performance air-oxygen mixer to form a mix of oxygen and air with a certain pressure.

Model: BR-MV14

SKU: MV14
£0.00Price
  •  

     

    Physical specifications

    Display

    10.4'' LCD Display

    Mode:

    pneumatically powered and electrically controlled system

    Working Mode:

    Closed; Semi-Closed; Semi-Open

    Trolley:

    Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions

    Technical Specifications:

    Gas requirement:

    Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa

    safety valve

    <12.5 kPa

    Respiratory Rate

    1bpm~99 bpm

    gas concentration output

    21% ~ 100% adjustable

    Maximum Minute ventilation

    ≥18L / min, permissible deviation: ±20 %

    Breathing Mode

    A/C, SIPPVIPPVSIMVPSIMVCPAPPCVPEEPSIGNSPONT, MANUAL

    Ventilator compliance

    : ≤30 Ml/kPa

    PEEP:

    0~1.0KPa

    Ptr:

    ,-1.0~1.0 kPa

    IMV frequency:

    1 ~ 12 times/min, permissible deviation: ±15%.

    I/E Ratio:

    4:1~1:4

    Tidal Volume range

    0~2000ml resolution ratio: 10ml

    Inspiratory Plateau:

    0~1s

    Maximum Safety Pressure

    ≤ 6.0kPa

    Sigh:

    one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point and the permissible error is -10% or 15%.

    Pressure limit range:

    1.0~ 6.0kPa

    Airway pressure alarm: Audible and visual and with yellow and red color indicating

    upper alarm: 0.9 kPa ~ 5.4 kPa

    lower: 0.5kPa

    Ventilation Volume alarm:

    ≥ 18L/ MIN

    Power Supply Alarm

    Ac/dc power supply are after failing to send out alarm immediately
    Alarm Time: keep >120s

    The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.

    Operating conditions

    Ambient temperature:

    10 ~ 40oC

    Relative humidity:

    no higher than 80 %

    Atmospheric pressure:

    860 hPa ~ 1060 hPa

    Power requirement:

    100-120 Vac, 50/60 Hz;

    Air source requirement:

    Medical oxygen and laughing gas with a rated pressure ranging from 0.3 to 0.5MPa.

    Attention: the AC power supply used for the anesthesia machine must be well grounded.

    Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System.

    Storage

    Ambient temperature:

    -15oC ~ +50oC

    Relative humidity:

    no higher than 95 %

    Atmospheric pressure:

    86 kPa ~ 106 kPa.

     

    It should be stored in a room without corrosive gas and well ventilated

    Package

    packaging box

    comply with the requirement of GB/T 15464

     

    Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation

     

    Moisture protection and rain protection to ensure that product is protected from natural damage.

    Safty & Alarm

    Gas source pressure

    Immediate alarm; “high level alarm” indicator flickers raising an audible alarm.

    Airway Pressure Alarm

    The airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The “high level alarm” indicator flickers raising an audible alarm.

    Tidal Volume Alarm

    Lower than 50ml, The alarm will be raised after one breathing cycle.

    Ventilation Volume Alarm

    Lower: 0~12L/Min; Up: 18L/ Min

    Battery Voltage Alarm

    It alrms when the battery voltage is lower than 10.5V.

    Power On Self Test

    Self test before working, assure it's working normally.

     