ADULT · PEDIATRIC
Application
The ventilator is suitable for various kinds of medical institutions for cardiopulmonary resuscitation respiratory support. Acute respiratory failure caused by various reasons or incomplete oxygenation dysfunction. Indra-operation, post-operation respiratory support, other respiratory treatment.
BR-MV15A ICU Ventilator
Features
l 15” TFT LCD touch screen displays the ventilation parameters, alarm information and waveforms.
l Air compressor runs stably with low noise, creates quiet environment for doctors and patients.
l Integration breathing circuit design.
l Multiple working modes such as volume control and pressure limit.
l Multiple parameters monitoring interface to show clear parameter to the operator, allow operator know the patient conditions in all aspects.
l Real time pressure-time, flow-time loop graphics, pressure volume loop graphics and high precision ETCO2, O2 concentration detection function included.
l Electrical PEEP control
l Breathing valve is disassembly and can be disinfected.
Safety
l Three levels alarm system, visual and sound alarm information.
l With multiple type of alarms, reminding and protection functions.
l Self-check before running, eliminate system mulfuction.
l Advanced power management control technology.
l With built-in backup power supply, when normal power supply shut down, back-up power supply will start work automatically.
l Separate design of electric and gas, keep safety running of ventilator.
Specifications
Ventilation modes:
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV
Ventilator parameter ranges
Tidal volume(Vt)
0 ~ 2000mL
Frequency (Freq)
Oxygen concentration
1bpm ~ 100 bpm
21%-100%
I/E
4:1~1:8
PEEP
0cmH2O ~ 40 cmH2O
Pressure Limit
20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequency (Freq)
0 /min ~ 100 /min
Tidal volume(Vt)
0 mL ~ 2500 mL
MV
Airway pressure
Dynamic lung compliance testing
ETCO2 Concentration
0 L/min ~ 99 L/min
0 cmH2O ~ 100 cmH2O
1mL/cmH2O ~ 1000mL/cmH2O
0mmHg ~ 152mmHg (0%-20%)
Oxygen concentration
15 % ~ 100 %
Oscillographs display:
P-T(pressure – time)
F-T(flow – time)
P-V loop (pressure – volume loop)
V-T(volume – time )
ETCO2-T (End-tidal CO2 - time)
Alarms and protection
The AC power failure alarm
Power failure or no connection
Internal battery backup low voltage alarm
≤11.3±0.3V
No tidal volume
No tidal volume within 6s
High Minute Volume alarm
Low Minute Volume alarm
5L/min-99L/min
1L/min-30L/min
High Airway pressure alarm
Low Airway pressure alarm
High oxygen concentration alarm
Low oxygen concentration alarm
20cmH2O-100cmH2O
0cmH2O-20cmH2O
19%-100%
18%-99%
Continuous pressure alarm
(PEEP+1.5cmH2O) over 16s
Suffocation warning
Fan error
Oxygen deficit
5-60s
Show on screen
Show on screen
The maximum limited pressure
<12.5 kPa
Working conditions
Gas source
O2, AIR
Pressure
280kPa-600kPa
Voltage
~220V±22V
Power frequency
50Hz ± 1Hz
Input power
900VA (with air compressor)
250VA (without air compressor)
Ventilation modes:
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV
Ventilator parameter ranges
Tidal volume(Vt)
0 ~ 2000mL
Frequence (Freq)
Oxygen concentration
1bpm ~ 100 bpm
21%-100%
I/E
4:1~1:8
PEEP
0cmH2O ~ 40 cmH2O
Pressure Limit
20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequence (Freq)
0 /min ~ 100 /min
Tidal volume(Vt)
0 mL ~ 2500 mL
MV
Airway pressure
Dynamic lung compliance testing
ETCO2 Concentration
0 L/min ~ 99 L/min
0 cmH2O ~ 100 cmH2O
1mL/cmH2O ~ 1000mL/cmH2O
0mmHg ~ 152mmHg (0%-20%)
Oxygen concentration
15 % ~ 100 %
Graphical display:
P-T(pressure – time)
F-T(flow – time)
P-V loop (pressure – volume loop)
V-T(volume – time )
ETCO2-T (End-tidal CO2 - time)
Alarm and protection
The AC power failure alarm
Power failure or no connection
Internal battery backup low voltage alarm
≤11.3±0.3V
No tidal volume
No tidal volume within 6s
High Minute Volume alarm
Low Minute Volume alarm
5L/min-99L/min
1L/min-30L/min
High Airway pressure alarm
Low Airway pressure alarm
High oxygen concentration alarm
Low oxygen concentration alarm
20cmH2O-100cmH2O
0cmH2O-20cmH2O
19%-100%
18%-99%
Continuous pressure alarm
(PEEP+1.5cmH2O) over 16s
Suffocation warning
Fan error
Oxygen deficit
5-60s
Show on screen
Show on screen
The maximum limited pressure
<12.5 kPa
Working conditions
Gas source
O2,AIR
Pressure
280kPa-600kPa
Voltage
~220V ±22V
Power frequency
50Hz ± 1Hz
Input power
900VA (with air compressor)
250VA (without air compreesor)