top of page

Product Application:

CPAP system is used for newborn baby with stronger spontaneous breath and slightly respiratory failure, and applies to pediatric clinics for NCPAP oxygen uptake

 

Product Performance and features :

1. CPAP ventilation mode keeps children spontaneous breath and reduces children breathing work, continues to provide CPAP mode of constant flow and

limited pressure for newborn baby;

2. Delivery air/O2 mixture with properly adjusted O2 concentration and continuous positive pressure to premature baby, newborn baby and child, which will

highly improve the safety and efficacy of oxygen therapy;

3. Equipped with built-in oxygen/air blender, ensuring O2 concentration adjustment to be stable ;

4. High quality oxygen sensor can monitor real-time O2 concentration , alarm over limitation, ensure safety of breathing O2;

5. Reliable CPAP control valve and advanced pressure monitoring system improve CPAP adjustment precision.

6. 3’’ display

 

Product Working condition

l Air source O2AIR

l Air source pressure 280 kPa 600 kPa

l Power supply 220 V±22 V, 50 HZ with air compressor

100-240V,50/60HZ(without air compressor

l Input power 900VAwith air compressor

250VA without air compressor

 

 

RC-BV200C CPAP System

£0.00Price
  • Ventilation Mode l CPAP

    Main technical parameters

    Adjusting Parameter

    l Inhaled oxygen concentration 21 %100 %

    l CPAP 0cmH2O15 cmH2O

    l Continuous flow 2L/min20L/min

     

    Monitoring Parameter

    l Inhaled oxygen concentration 15 % 100 %

    l CPAP 0 cmH2O 60 cmH2O

    l Continuous flow 2 L/min 20 L/min

    Alarm and protection

    l The AC power cut-off alarm

    l Low air (oxygen) input pressure alarm

    l Air / oxygen input unbalanced pressure alarm

    l High oxygen concentration alarm

    l low oxygen concentration alarm

    l Pressure limit

Lepu Medical Covid-19 AntiBody test Leaflet
Lepu Medical Covid-19 AntiBody test Instructions
bottom of page