Product Features:

Upright, RC-V2000B3 ventilator is a multi-functional ventilator. Suitable to be used in departments of internal medicine, surgery, Obstetrician/Gynecology, pediatrics, emergency room and ICU. Including IPPV, SIPPV, SIMV, SIGH, MANUAL, PEEP, IMV, VCV, PCV, PSV, CPAP, Psupp etc. functions and monitoring O2 concentration. High-resolution 10.4″ TFT LCD, real-time parameters display, such as pressure/time and flow-speed/time waves etc. Including cart, humidifier, mechanical arm , compressor and manual ventilator.

 

Product Technical Data:

■ Display mode

HD 10.4 TFT color LCD display

■ Ventilation mode

IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, (IMV, VCV, PCV,

PSV, CPAP)

■ ventilation parameters

Tidal volume of 50 ~ 1500 ml (20 ~ 1500 ml)

Frequency 1 ~ 99 times / min

SIMV frequency 1 ~ 20 times / min

Absorption ratio of 2:1 to 1:8 call (4:1 ~ 1:8)

Inspiratory trigger pressure of -1.0 ~ 2.0 kPa

Flow trigger sensitivity of 1 ~ 8L/min

PEEP 0 ~ 2.0 kPa

1.0 ~ 6.0 kPa pressure control

Support 1 ~ 4 kPa pressure

Sigh inspiratory time of 1.5 times (70 to 150 adjustable)

Inspiratory Ping Taiwan 0 ~ 50%

Oxygen concentration of 21% to 100%

■ Ventilation monitoring

Tidal volume, ventilation, IPPV frequency, SIMV frequency, total respiratory rate, ratio, peak airway pressure, pressure - time waveform, flow rate - the time wave, positive end expiratory pressure, inspiratory trigger pressure,

Ping suction units, oxygen concentration (autonomous breathing frequency, lung compliance, airway resistance, flow rate - volume loops, pressure - volume loops)

■ Security Alarm System

Oxygen concentration range of 21% upper limit set to 100%

Limit setting range of 10% to 80%

Airway pressure alarm limit setting range of 0.3 ~ 6.0 kPa

Lower limit range of 0.2 ~ 5.0 kPa

Minute ventilation alarm limit setting range 3.0 ~ 30 L / min

Limit setting range 1.0 ~ 10 L / min

Respiratory rate alarm limit setting range of 50 to 99 times / min

Lower limit range of 0 to 50 times / min

Maximum tidal volume alarm setting range 0 ~ 2000ml

Lower limit range of 0 ~ 1800ml

High pressure continued to alarm when the alarm has been higher than 2.5 kPa

Suffocation alarm input 15s without tidal volume, the sound, light alarm

Gas source undervoltage alarm

Power failure alarm

Intubation off alarm

■ Power Supply AC 220V ± 10% 50Hz ± 1Hz

UPS power supply, battery

■ arm

■ (air compressor)

Department of matching items in brackets

RC-V2000B3