Physical specifications 
Screen: 5.4'' Display Screen
Suitable Adult & Child
Mode: pneumatically powered and electrically controlled system
Working Mode: Closed; Semi-Closed; Semi-Open
Circuit Breathing circuit integrated standards
Tube: 4Tubes Flowmeters:  O2:0.1~10L/Min,  N2O:0.1~10L/Min
Trolley: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Drawer unit ONE drawer totally extractable

BR-AM05 Anesthesia Machine

£0.00Price
  • Gas pressure: Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ;   NO2: 0.32 MPa to 0.6 MPa.
    safety valve ≦12.5  kPa
    Respiratory Rate 1~99bpm
    Flow meter O2:0.1~10L/Min,  N2O:0.1~10L/Min
    oxygen concentration in the mixed gas N2O/O2 > 21%
    Rapid Oxygen Flush: 25~75 L/min
    modes of ventilation IPPV, SIPPV, IMV, SIMV, MANUAL
    I/E Ratio: 4:1 ~ 1:10
    Tidal Volume 0~1500ML
    Inspiratory trigger pressure -0.1kPa~1.0kPa
    Inspiratory Plateau: 0~1s
    Maximum safety pressure: ≤ 12.5 kPa
    Pressure limit range: 0 ~ 6.0  kPa
    Airway pressure alarm: Audible and visual and with yellow and red color indicating: Lower: 0.2kPa~5.0kPa;   Upper: 0.3~6.0  kPa
    ±0.2 kPa
    Power Supply Alarm "Ac/dc power supply are after failing to send out alarm immediately
    Alarm Time: keep >120s"
    The airway pressure continues to be higher than 15 hPa?±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.