Application:
The ventilator is suitable for various kinds of medical institutions for cardiopulmonary resuscitation respiratory support. Acute respiratory failure caused by various reasons or incomplete oxygenation dysfunction. Intra-operation, post-operation respiratory support, other respiratory treatment.
Simplicity: easy to use, easy to move with 4 wheels.
Choice: adapt the equipment to your patients and procedures freely
Patient Centered Ventilation: Precision in an anesthesia ventilator, from conventional ventilation to advanced modes, including 8 modes: IPPV; A/C; PCV; SIMV; SPONT/CPAP;PSV;SIGH;MANU.
Designed and manufactured by Perlong Medical with over 23 years experience in this area.
Flexible configurations to suit your needs.
International standard and advanced technology suitable for wide range use.
Compact interface and big screen give you better operating experience.
Over 2,000 units installed in the world.
BR-MV15 ICU Ventilator
Features
12.1” TFT LCD screen displays the Ventilation parameters, Alarming information and Waveform.
Recycle breathing pipe, ensure easy operating and keep tidy.
Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
Vaporizer with temperature, flow compensation and self-lock function, keep safety anytime.
Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects;
Real time pressure-time, flow-time loop graphics and high precision O2 concentration detection function included.
Stable and low noise air compressor, create quiet work environment for doctors.
Safety
Three level alarming system, visual and sound alarm information.
With lots of alarming, reminding and protection functions.
Advanced power management control technology.
With built-in backup power source, when outside power source goes off, back-up power source starts to work.
Self-check before running, eliminate system mistake.
Separate design of electric and gas, keep safety running of ventilator.
Ventilation modes:
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV,SIGH,MANU
Ventilator parameter ranges
Tidal volume(Vt)
0 ~ 2000mL
Frequency (Freq)
Oxygen concentration
1bpm ~ 100 bpm
21%-100%
I/E
4:1~1:8
PEEP
0cmH2O ~ 40 cmH2O
Pressure Limit
20 cmH2O ~ 100 cmH2O
Monitored Parameters
Frequency (Freq)
0 /min ~ 100 /min
Tidal volume(Vt)
0 mL ~ 2500 mL
Oxygen concentration
15 % ~ 100 %
Oscillographs display:
P-T(pressure – time)
F-T(flow – time)
P-V loop (pressure – volume loop)
Size
1st Wooden case packing size(main engine):L560*W 560*H 605mm , GW:35KG ; NW:17KG
2nd Wooden case packing size(air compressor):L670*W700*H 1160 mm , GW:79KG ; NW:46.2KG
Alarm and protection
The AC power failure alarm
Power failure or no connection
Internal battery backup low voltage alarm
≤11.3±0.3V
No tidal volume
No tidal volume within 6s
High Minute Volume alarm
LowMinute Volume alarm
5L/min-99L/min
1L/min-30L/min
High Airway pressure alarm
Low Airway pressure alarm
High oxygen concentration alarm
Low oxygen concentration alarm
20cmH2O-100cmH2O
0cmH2O-20cmH2O
19%-100%
18%-99%
Continuous pressure alarm
(PEEP+1.5cmH2O) over 16s
Suffocation warning
Fan error
Oxygen deficit
5-60s
Show on screen
Show on screen
The maximum limited pressure
<12.5 kPa
Working conditions
Gas source
O2,AIR
Pressure
280kPa-600kPa
Voltage
-220V±22V
Power frequency
50Hz±1Hz
Input power
900VA(with air compressor)
250VA(without air compressor)